Patient Monitor

Patient Monitor ANALOGIC ASC553Specification :

  • Display mode                    12,1” color TFT

Display of the waveform of patient’s ECG, Respiration and SPo2.

Complete ECG monitoring through selection by different leads positioning.

Blood pressure may be measured in the mode of adult, infant, and neonate, which may be selected via menu.

  • Power supply                    220V/50Hz.
  • Power Consumption         80 Watt
  • Alarm mode                      Audible visual-alarm
  • Gross weight                     12 Kg
  • Dimension                         500(L) x 320(W) x 460 (H) mm

Parameter :

ECG/RESP/TEMP

Input Impedance                                 > 5 MΏ

Input linearity                                       ± 3%

ECG Gain                                            ½                     1 grid/mV ±10%

1                                            2 grid/mV ±10%

2                                            4 grid/mV ±10%

Auto                 Automatic gain control

Respiration Waveform Gain               4 options available (1/2, 1, 2 and 4)

Lead selection                                     GND, Standard I, II, & III

Limb purpose lead AVR, AVL, & AVF

ECG Chart Speed                               12.5 mm/s, 25 mm/s, 50 mm/s

Respiratory monitoring                       Range 0 – 100 bpm

Accuracy ± 10%

Record speed 6.25 mm/s, and 12.5 mm/s

TEMP monitoring                                Range 28.0o C – 43.2oC

Accuracy 0.2oC

Response Time < 2 min

Non-invasive Blood Pressure

Measuring technology                         Automatic measurement by intelegent Oscillting method.

Cuff inflation time                                < 30 second (Standard adult cuff 0 to 250 mmHg)

Deflation time at measurement          < 90 second on the average

Initial cuff inflation pressure                Adult mode                  180 mmHg

Infant mode                 100 mmHg

Neonate mode            80 mmHg

Max inflation pressure allowed            300 mmHg (when greater than this value safetu valve will release the air automatically)

Default measuring mode                    Manual, adult mode

Measuring interval                               0 – 250 minutes adjustable

Pulse Oxygen Saturation (SPo2)

Sensor                                                Dual-wavelength LED

SpO2 35% ~ 100%

Resolution of SpO2 1 %

Accuracy of SpO2 90 – 99%                     ±2%

70 – 89%                     ±4%

40 – 69%                     ±6%

Pulse Range                                       30 ~ 240 bpm

Pulse accuracy                                   ±2 bpm

Sampling frequency of SpO2 Once a second